5.11 Setting the 8954 VOLTAGE/TEMP UNIT
────────────────────────────────────────────────────
WARNING
A common GND is used for voltage and temperature input on
each channel. When using the voltage and temperature inputs
simultaneously, do not connect them both at the same time, as
this may damage the object you are measuring. (You cannot
measure the voltage and temperature simultaneously on the
same channel.)
5.11.1 Setting Voltage Measurement
Flashing cursor
Function
display
Meaning
:
Move the cursor up in the selection
window.
:
Move the cursor down in the selection
window.
5.11.2 Setting Temperature Measurement
Flashing cursor
Thermocouple Measurement input range Thermocouple Measurement input range
K -200 to 1350 N -200 to 1300
E -200 to 800 R 0 to 1700
J -200 to 1100 S 0 to 1700
T -200 to 400 B 300 to 1800
W 0 to 2000
5.11 Setting the 8954 VOLTAGE/TEMP UNIT
Procedure
Screen: LIST, ONE CH (CHANNEL)
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Mode
item to be
set.
3. Use the function keys to select
VOLTAGE
.
For details on other common settings, see Quick Start Section 6.3.
Procedure
Screen: LIST, ONE CH (CHANNEL)
(1) Setting the measurement mode
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the
Mode
item to be
set.
3. Select the temperature range and thermocouple
to be used with the function keys.