5.10 8953-10 HIGH RESOLUTION UNIT
────────────────────────────────────────────────────
Flashing cursor
Function
display
Meaning
:
Anti-aliasing filter is disabled.
:
Anti-aliasing filter is enabled.
NOTE
5.10 8953-10 HIGH RESOLUTION UNIT
The 8953-10 HIGH RESOLUTION UNIT contains an internal anti-aliasing
filter required for FFT analysis. Enable the anti-aliasing filter when doing
FFT analysis.
For details on other common settings, see Quick Start Section 6.3.
Procedure
Screen: LIST, ONE CH (CHANNEL)
1. Use the Menu keys to display the desired screen.
If the List screen is displayed, press the
F3 (
PAGE)
key to display the OPTION screen.
2. Move the flashing cursor to the
A.A.F
item to be
set.
3. Use the function keys to make a setting.
We recommend using an input module equipped with an anti-aliasing filter
that can be enabled to minimize sampling distortions during FFT analysis.
Refer to Appendix 3.9, "FFT Function" for more information about aliasing
distortion and anti-aliasing filters.
Refer to Section 4.3.3, "Setting the Frequency Range" for details about the
relationship between the anti-aliasing filter cutoff frequency and the
frequency range and time axis range.
When using the recorder function, the anti-aliasing filter cannot be used even
if it is enabled.
When using the 8953, the anti-aliasing filter cannot be used even if it is
enabled.