31
────────────────────────────────────────────────────
External Memory Devices
────────────────────────────────────────────────────
Press the function ke y corresponding to the setting
displayed in the screen.
Move to the item to be set.
(5) Enter a name (file name) for the data to be saved.
Move the flashing cursor to "Save Name." Press
the input
F1 key to display the character input
screen. Enter a file name.
Regarding the input method, refer to "5.3.3
Character Entry Procedure" in the 8855 Instruction
Manual.
F1
input
Character input screen
(6) Specify the method of processing to be executed if
the entered file name already exists.
Move the flashing cursor to "Same Name," and
select a processing method.
Auto Automatically adds a number after the
entered file name
Overwrite Deletes the old file and saves the new
file
Error Displays an error message and stops
saving the data
(7) When saving data in text form, set a data thinning
rate.
Move the flashing cursor to "Save Thin," and select
a thinning rate.
(8) Set the channel to be saved.
Move the flashing cursor to the position of the
channel to be saved. For an analog channel, make
settings individually for each unit. For a logic
channel, make all settings at once.
(9) Save data.
Press the exec.
F9 key to save the data.
To cancel the data saving command, press the
cancel
F10 key.
F9
exec.
F10
cancel