Quick Start Guide

30
────────────────────────────────────────────────────
External Memory Devices
────────────────────────────────────────────────────
Storage of measurement data
Displays the
information of the
file to be saved.
Set the processing to
be executed if the
entered file name
already exists.
Specify the range
of data to be
saved.
Displays the
execution
command.
Select a file save
format.
Enter a name for
the file to be
saved.
When data is
saved as text, set
a thinning rate.
Specify the
channel to be
saved.
(1) Display the file screen, and select a data recording
media (refer to "Selection of recording media").
(2) Press the SAVE F2 key.
Press the function ke y corresponding to the setting
displayed in the screen.
Move to the item
to be set.
(3) Select a file save format.
Using the cursor keys, move the flashing cursor to
"Save Type," and select a save setting using the
function key.
F2
SAVE
Setting Stores setting conditions
Wave binary Stores waveform data in binary format
(for 8855)
Wave text Stores waveform data in text format
(for Excel). Data stored in text format
cannot be loaded into the 8855.
(4) Specify a saving range.
Move the flashing cursor to "Save Area," and
specify the range of data to be saved.
When A-B cursors are used, specify the range to
be saved.
Select either "ALL WAVE" or "A-B W AVE."