Quick Start Guide
27
────────────────────────────────────────────────────
Measurement Method
────────────────────────────────────────────────────
(3) A-B cursors
A-B cursors are used to read frequency and numerical values
such as maximum values. When the A.B CSR key is pressed,
the red LED lights up and the GUI for A-B cursor settings
appears. Select the type of cursor, ON/OFF of A-B cursors, and
movement for reading a numerical value.
< Key point >
The waveform display screen can be switched between 30 DIV
and 20 DIV by pressing the DISP
key.
The aforementioned allows easier reading, since values such
as those read by the cursors do not overlap with a waveform or
grid.
To exit from the A-B cursor setting mode, press the A.B CSR
key again, or press the W AVE, VALUE, or ESC key.
Cursor A
Cursor type
Vertical/horizontal cursor
Trace cursor
Selection of cursor display
A cursor only
(Gauge cursor possible)
cursor B
Selection of cursor m ov ement
The F7, F8, and A-B cursors move
when their toggle keys are marked.
Selection of cursor reading value display
Setting of ON/OFF of read value display