Measurement Method
────────────────────────────────────────────────────
3. Trigger setting
The following describes the procedure for
setting trigger conditions.
When there is a change from 0 V (0 G), the
trigger is activated and waveform acquisition
starts.
(1) Press the TRIG key to display the trigger s creen.
(2) Using the cursor keys, move the flashing cursor to
CH1 "Kind," and press the function key to change
the "Kind" to "LEVEL."
(3) Move the flashing cursor to CH1 "Parameter," and
set the level to "200 mV" and the slope to "
."
(4) Move the flashing cursor to "Trigger Mode," and set
the trigger mode to "SINGLE."
(5) Move the flashing cursor to "Pre-Trigger," and set
the pre-trigger to "5%."
Move to "Kind," "Parameter,"
"Trigger Mode," and "Pre-Trigger."
Press the function ke y corresponding to the setting
displayed in the screen.
Trigger screen
Set to level, 200 mV, and
Display the trigger screen.
Set to "Single" and "5%."