Data Sheet

9
VOLTAGE/TEMP UNIT 8954
(Accuracy at 23 ±5˚C (73 ±9˚F) and 30 to 80% relative humidity when zero-
adjust is performed after a 60-minute warm-up; accuracy guaranteed for 1 year)
Measurement functions
Number of channels: 1 voltage or temperature measurement channel
Input
Voltage input: isolated BNC connector, thermocouple input: Plug-in
terminal, Max. rated voltage to earth: 370 V AC, DC
(with input isolated
from the unit, the maximum voltage that can be applied between input channels
and chassis, or across input pins, without causing damage)
Voltage measurement
range
500 µV to 2 V/div, 12 settings, full-scale = 20 div,
low-pass filter: 1 Hz/5 Hz/50 Hz/500 Hz, the measurement resolution:
1/1600 of range
(using 16-bit AD)
Temperature
measurement range
10 ˚C, 100 ˚C/div, 2 settings, full-scale = 20 div,
low-pass filter: 1 Hz/5 Hz/50 Hz/500 Hz, the measurement resolution:
1/1000 of range
(using 16-bit AD)
Thermocouple range
K: -200 to 1350 ˚C, E: -200 to 800 ˚C, J: -200 to 1100 ˚C,
T: -200 to 400 ˚C, N: -200 to 1300 ˚C, R: 0 to 1700 ˚C,
S: 0 to 1700 ˚C, B: 300 to 1800 ˚C, W: 0 to 2000 ˚C
Reference junction compensation: internal/external
(interchangeable)
Maximum sampling rate
Voltage input: 100 kS/s,
temperature measurement: 4kS/s
(data updates every 250 µs)
Accuracy
Voltage input: DC amplitude: ±0.2 % f.s. Zero-position: ±0.2 % f.s.
Temperature measurement (K, E, J, T, N): ±0.1 % f.s. ±1 °C, ±0.1 % f.s. ±2
°C
(-200 °C to 0 °C)
, (R, S, W): ±0.1 % f.s. ±3 °C, (B): ±0.1 % f.s. ±4 °C
(400 to 1800 °C)
Reference contact compensation accuracy: ±0.1 % f.s. ±1.5 °C
(for reference
junction internal compensation)
Frequency
characteristics
Voltage input: DC to 20 kHz +1/-3 dB
Temperature measurement: DC to 1 kHz +1/-3 dB
Input resistance and capacitance
Voltage input: 1 MΩ, 60 pF
(when C is 10 kHz)
Temperature measurement: 4.8 MΩ or more
Max. allowable input
30 Vrms or 60 V DC
(the maximum voltage that can be applied between input
pins without causing damage)
F/V UNIT 8955
(Accuracy at 23 ±5˚C (73 ±9˚F) and 30 to 80% relative humidity when zero-adjust is
performed after a 30-minute warm-up; accuracy guaranteed for 1 year)
Measurement functions
Number of channels: 1, Measurements: frequency of voltage input,
power frequency, revolution speed, integration, pulse duty comparison,
pulse width
Input
Metal BNC connector, Max. rated voltage to earth: 30 Vrms or 60 V
DC
(with input isolated from the unit, the maximum voltage that can be applied
between input channels and chassis, or across input pins, without causing damage)
Measurement range
Frequency: 0.1 Hz to 5 kHz/div between DC to 100 kHz, 10 settings
Rotations: 10 (r/min) to 500 (r/min)/div, 4 settings
Power frequency: 50 Hz
(40 to 60 Hz)
, 60 Hz
(50 to 70 Hz)
Integration: 2 kcounts to 1 Mcounts/div between DC to 90 kHz
Pulse duty ratio: 100 % f.s. between 10 Hz to 100 kHz, 1 setting
Pulse width: 500 µs to 100 ms/div between 2.5 µs to 2 sec, 6 settings
Max. allowable input: 30 Vrms or 60 V DC
(the maximum voltage that can
be applied between input pins without causing damage)
,
full-scale = 20 div, low-pass filter: 5 Hz/500 Hz/5 kHz/100 kHz/OFF
Measurement resolution
Frequency, power frequency, rotations, pulse width, duty ratio: 1/200
of measurement range
Pulse duty ratio: 1/2000 of measurement range
Response time
10 µs + 50 ns or less
(when frequency is less than 300 Hz, measuring integration
or pulse width)
50 µs + 50 ns or less
(when frequency is less than 300 Hz, measuring pulse duty
ratio, rotation)
Accuracy
Frequency: ±0.1 % f.s.
(for settings other than 100k Hz f.s.), ±0.7 % f.s. (100
kHz f.s. setting)
Rotation: ±0.1 % f.s.
Power frequency: ±0.032 Hz
Pulse duty ratio: ±1 % f.s.
(10 Hz to 10 kHz)
, ±4 % f.s.
(10 kHz to 100 kHz)
Pulse width: ±0.1% f.s.
Other functions
Input pull up resistance: ON/OFF
(10 kΩ connected to 5 V)
Threshold level: -10 to +10 V variable, slope: rises and drops
Level: High, Low, Hold ON/OFF
Other functions
DIFFERENTIAL PROBE 9322 Power
(the POWER CORD 9328 is required
for connection)
LOGIC PROBE 9320-01/9327
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh; accuracy guaranteed for 1 year)
Function
Detection of voltage signal or relay contact signal for High/Low state recording
Input
4 channels (common ground between unit and channels), digital/contact input,
switchable
(contact input can detect open-collector signals), input impedance:
1MΩ
(with digital input, 0 to +5V), 500kΩ or more (with digital input, +5 to
+50V)
, pull-up resistance: 2kΩ (contact input: internally pulled up to +5V)
Digital input threshold
1.4V/2.5V/4.0V
Contact input
detection resistance
1.5kΩ or higher (open) and 500Ω or lower (short), 3.5kΩ or higher (open)
and 1.5kΩ or lower
(short), 25kΩ or higher (open) and 8kΩ or lower (short)
Response speed
9320-01: 500ns or lower, 9327: detectable pulse width 100ns or higher
Max. allowable input
0 to +50V DC (the maximum voltage that can be applied across input pins without damage)
LOGIC PROBE 9321-01
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh; accuracy guaranteed for 1 year)
Function
Detection of AC or DC relay drive signal for High/Low state recording
Can also be used for power line interruption detection
Input
4 channels (isolated between unit and channels), HIGH/LOW range switching
Input impedance: 100kΩ or higher
(HIGH range), 30kΩ or higher (LOW range)
Output (H)
detection
170 to 250V AC, ±DC (70 to 250V ) (HIGH range)
60 to 150V AC, ±DC (20 to 150V) (LOW range)
Output (L)
detection
0 to 30V AC, ±DC (0 to 43V) (HIGH range)
0 to 10V AC, ±DC (0 to 15V) (LOW range)
Response time
Rising edge 1ms max., falling edge 3ms max. (with HIGH range at 200V
DC, LOW range at 100V DC)
Maximum allowable
input voltage
250Vrms (HIGH range), 150Vrms (LOW range) (the maximum voltage that can
be applied across input pins without damage)
DIFFERENTIAL PROBE 9322
(Accuracy at 23 ±5°C/73 ±9°F, 35 to 80% rh, after 30 minutes of
warm-up time; accuracy guaranteed for 1 year)
Function
For high-voltage floating measurement, power line surge noise
detection, RMS rectified output measurement
DC mode
For waveform monitor output, frequency characteristics: DC to 10MHz
(±3dB), amplitude accuracy: ±1% of full scale (at max. 1000V DC), ±3%
of full scale
(at max. 2000V DC) (full scale: 2000V DC)
AC mode
For detection of power line surge noise, frequency characteristics: 1kHz to 10MHz ±3dB
RMS mode
DC/AC voltage RMS output detection, frequency characteristics: DC, 40Hz
to 100k Hz, response speed: 200ms or less
(400V AC), accuracy: ±1% of full
scale
(DC, 40Hz to 1kHz), ±4% of full scale (1kHz to 100kHz) (full scale: 1000V AC)
Input
Input type: balanced differential input, input impedance/capacitance:
H-L 9MΩ/10pF, H/L-unit 4.5MΩ/20pF, Max. rated voltage to earth:
when using grabber clip 1500V AC/DC
(CAT II ), 600V AC/DC (CAT III),
when using alligator clip: 1000V AC/DC
(CAT II), 600V AC/DC (CAT III)
Maximum allowable input voltage
2000V DC, 1000V AC (CAT II), 600V AC/DC (CAT III)
Output
Voltage divider for 1/1000 of input, BNC connectors (output switchable for 3 modes DC, AC, RMS)
Power source
Power terminal of the input units, or use with AC ADAPTER 9418-15
(DC 12V)
Dimensions and mass: approx. 104.7(4.1 in)W × 28(1.1 in)H × 164.5(6.5 in)D mm
approx. 160 g (5.6 oz) Supplied accessories: None
Dimensions and mass: approx. 104.7(4.1 in)W × 28(1.1 in)H × 164.5(6.5 in)D mm
approx. 150 g (5.3 oz) Supplied accessories: None
Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 30 cm
(0.98 ft), approx. 150 g (5.3 oz)
Note: The unit-side plug of the 9320-01 and 9327 is different from the 9320.
Cable length and mass: Main unit cable 1.5 m (4.92 ft), input section cable 1 m (3.28 ft),
approx. 320 g (11.3 oz) Note: The unit-side plug of the 9321-01 is different from the 9321.
Cable length and mass: Main unit cable 1.3 m (4.27 ft), input section cable 46 cm
(1.51 ft), approx. 350 g (12.3 oz)
WAVE PROCESSOR 9335
Distribution media
One CD-R
Operating
environment
Computer equipped with Pentium
(133 MHz)
or better CPU and at least 32
MB of memory, and running under Windows 95/98/Me, Windows NT
4.0/2000/XP, or Windows Vista 32-bit type
(recommended system: Pentium
(200 MHz) or better with at least 64 MB of memory)
Display functions
Waveform display/X-Y display/digital value display/cursor function/
scroll function/maximum number of channels
(32 channels analog, 32
channels logic)
/gauge display
(time, voltage axes)
/graphical display
File loading
Readable data formats (.MEM, .REC, .RMS, .POW)
Maximum loadable file size: Maximum file size that can be saved by a
given device
(file size may be limited depending on the computer configuration)
Data conversion
Conversion to CSV format, tab delimited, space delimited/data culling
(simple)
/convert for specified channel/batch conversion of multiple files
Print functions
Print formatting
(1 up, 2-to-16 up, 2-to-16 rows, X-Y 1-to-4 up)
/preview/hard
copy functions usable on any printer supported by operating system
Other
Parameter calculation/search/clipboard copy/launching of other
applications
LAN COMMUNICATOR 9333
Distribution media
One CD-R
Operating
environment
Computer equipped with Pentium (133 MHz) or better CPU, running
under Windows 95/98/Me or Windows NT 4.0/ 2000/Xp operating
system, with network adapter installed and configured to use TCP/IP
protocol, and at least 64 MB of memory.
HiCORDER side
Standard LAN connector
Communications
Ethernet, TCP/IP
Remote control
Remote control of MEMORY HiCORDER (by sending key codes and receiving
images on screen), print repor ts, print images from the screen, receive
waveform data in same format as waveform files from the MEMORY
HiCORDER (binary only)
Waveform data
acquisition
Accept auto-saves from the MEMORY HiCORDER, same format
as auto-save files of MEMORY HiCORDER (binary only), print
automatically with a MEMORY HiCORDER from a PC. The
MEMORY HiCORDER's print key launches printouts on the PC
Waveform viewer
Simple display of waveform files, conversion to CSV format, Scroll
function, enlarge/reduce display, display CH settings.