User Manual

70
────────────────────────────────────────────────────
5.1 Outline of the Recorder Function
────────────────────────────────────────────────────
5
.1 Outline of the Recorder Function
The recorder function has the following features.
(1) Real time display and printing of the input signal
(2) Real time continuous recording of the input signal
(3) All input channel data are recorded on the same time axis.
Since data for all channels can be superimposed, the relative relationship
between input signals can be observed visually.
(4) Time axis setting
20 ms/DIV to 1 h/DIV
(5) Time axis resolution 100 points/DIV (printer)
(6) Sampling period
1, 10, 100μs, 1, 10, 100 ms
Can be selected, from 1/100 of the time axis setting
(7) Memory capacity
16 M words: 2000 DIV
64 M words: 10000 DIV
Arbitrarily (Set from 1 division to the maximum number of divisions at 1-
division intervals)
(8) Waveform magnification/compression display and printout
Time axis direction: ×1to×1/500
Voltage axis direction: ×10 to ×1/2 (single, X-Y single)
×5to×1/4 (dual to hex, X-Y quad)
With the variable function, vernier function
(9) Display format
Time axis waveform: single, dual, quad screen display (LCD), oct, hex screen
displayprinter only
X-Y waveform: dot, line, singlequad screen
Time axis waveform: single, dual, quad screen display
(10) Scrollable display
The most recent 2000 (64 M words: 10000) divisions of the data are stored in
memory.
It is possible to scroll back for easy review.
(11) Additional recording function
The first set of measurement data is preserved, and recording of the second
set of measurement data starts after the first set.
(12) Logging function
Numeric printout of waveform data.