59
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
────────────────────────────────────────────────────
4
.3.12 Setting the Averaging Function
Screen: STATUS
1
Press the STATUS key to call up the
STATUS1 screen.
Move the flashing cursor to the averaging
item, as shown in the figure on the left.
Use the JOG control or the function keys
to set the averaging count.
:Move the cursor up in the selectio
n
window.
:Move the cursor down in the
selection window.
Selection
Selection
Method
Method
After starting the measurement, the
averaging count and the current waveform
data count are shown on the screen.
NOTE
・The averaging function allows capturing several instances of a waveform
and determining the average.
・This makes it possible to eliminate noise and irregular signal components.
・The higher the number of averaging instances, the more effectively will
noise be suppressed.
・When the averaging function is used, logic waveform is not displayed.
・When the memory segmentation function is used, averaging is not available.
・Averaging and waveform processing cannot be carried out simultaneously.
・The averaged waveform becomes available for waveform processing when the
averaging setting is turned OFF following measurement.
・When the averaging function is used, the maximum recording length is
reduced to 25% of the normal value.