User Manual
58
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
────────────────────────────────────────────────────
4
.3.11 Setting the Overlay Function
Screen: STATUS
1
Press the STATUS key to call up the
STATUS1 screen.
Move the flashing cursor to the overlay
item, as shown in the figure on the left.
Use the function keys to make the
selection.
:Overlay is not performed.
:Overlay is performed.
Selection
Selection
Method
Method
NOTE
・Overlay is performed without clearing the currently displayed waveform (if
trigger mode is REPEAT or AUTO). This allows comparison to the immediately
preceding waveform.
・If trigger mode is SINGLE, measurement terminates after one set of data has
been collected. Therefore the overlay setting is invalid.
・While overlay is being executed, operations on the display screen (waveform
scrolling, change in time-axis magnification/compression ratio, change in
zero position, change in display size, and jumping to another display screen
using a VIEW function (see Section 11.6)) are disabled.
・When manual printing or trace cursor reading carried out, only the last
waveform will be done.
・Use Report Print to print overlaid waveforms. (see Section 13.6.7)
・If one of the following settings is changed, the overlay waveform display
terminates and only the last waveform is shown:
STATUS1: Display format, Display size
CHANNEL1: Input settings
・The overlay function cannot be set together with roll mode.