4.3 STATUS Settings (MEM)
────────────────────────────────────────────────────
Move the flashing cursor to type, and selec
t
the format.
Data are stored as binary data.
Data are stored as text data.
Data stored in the text format is
not readable by the 8826.
Selection
Selection
When the data format to be saved is set to
text data, the intermittent setting item is
displayed. Use the function keys or Jog
control to make a setting.
Move the cursor up in the selection
window.
Move the cursor down in the
selection window.
Selection
Selection
Set the file name. For the input method,
refer to Section 9.8.4.
When using auto-save, a number is
appended to the name you specify as the file
name. This then becomes the file name. If
you start procedures before specifying a file
name, AUTO is automatically enabled.
Do not use the characters and symbols below
in a file name. A file containing any of these
characters/symbols in its file name cannot be
processed under Windows 2000 or XP. Half-
size symbols: +, =, [,]
NOTE
When auto-save is enabled, the storage channel cannot be selected. Data for the
channel for which the waveform is displayed is saved.
For details on auto saving file name, refer to Section 14.7.
When both auto-print and auto-save are enabled, auto-save usually takes
precedence. However, if roll mode is enabled, auto-print will execute first.
For details on connected SCSI device, refer to Section 14.4.
When cursor A and B are enabled, partial saving is executed.
During automatic storage, if the STOP keyispressedtwicetointerrupt
measurement, waveforms taken prior to the interruption are stored
automatically.
The directory and the number of files that can be stored in the directory are
limited. For details, see Section 14.7.
When "COMMUNI" (communications) is selected as Media, Storage Method and
Type are not displayed.