User Manual
54
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
────────────────────────────────────────────────────
4
.3.8 Setting the Roll Mode
Screen: STATUS
1
Press the STATUS key to call up the
STATUS1 screen.
Move the flashing cursor to the roll mode
item, as shown in the figure on the left.
Use the function keys to make the
selection.
:Normal recording is carried out.
:The waveform is displayed
immediately at the start of
recording.
Selection
Selection
Method
Method
NOTE
・This mode can be used at a time axis range setting of 10 ms/DIV or slower.
・In normal recording, the waveform is displayed only after all data of the
recording length have been captured. At low sampling speed settings, this will
cause a considerable delay between the start of measurement and the
appearance of the waveform on the display.
・When roll mode is set to ON, the waveform is displayed immediately at the
start of recording (the screen scrolls).
・When the time axis range of 10 ms/DIV or faster is set, normal recording is
carried out even if roll mode is set to ON.
・Roll mode cannot be set together with the external sampling.
・When Roll Mode is set to ON, the settings for Overlay, Averaging,
Sequential Saving, Calculating Waveform data, and Waveform evaluation
are automatically turned OFF.
(When the display format is set to "X-Y," you can turn ON Overlay with Roll
Mode ON.)