User Manual
45
────────────────────────────────────────────────────
4.3 STATUS Settings (MEM)
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
1
0
1
1
1
2
1
3
1
4
A
4
.3.2 Setting the Time Axis Range
Screen: STATUS1, DISPLAY
Call up the STATUS1 or DISPLAY screen
.
Move the flashing cursor to the time/div
item, as shown in the figure on the left.
Use the JOG control, the function keys or
the TIME/DIV key to make the selection.
Method
Method-1
External sampling can be used to set「EXT.」
External sampling →See section 17.4.
Method
Method-2
Call up the STATUS1 or DISPLAY screen
.
Use the TIME/DIV key to make the
selection.
The TIME/DIV key can be used regardless
of where the flashing cursor is located.
:Use the TIME/DIV ke
y
・Set the speed for inputting and storing the waveform of the input signal.
・Time axis range setting expresses the time for 1 DIV.
・The sampling period is 1/100th of the set value for the time axis range. (100
samples/DIV)
Any number of samples per division can be set when EXTERNAL is selected
as the time axis range.