User Manual

40
────────────────────────────────────────────────────
4.1 Outline of the Memory Recorder Function
────────────────────────────────────────────────────
C
hannel 32 ch 16 ch 8ch 4ch
M
emory capacity 16 M words 5000 DIV 10000 DIV 20000 DIV 40000 DIV
M
emory capacity 64 M words 20000 DIV 40000 DIV 80000 DIV 160000 DIV
4
.1 Outline of the Memory Recorder Function
The memory recorder function has the following features.
(1) After being stored in the internal memory, input signal data can be displayed and
printed.
(2) All input channel data are recorded on the same time axis.
Since data for all channels can be superimposed, the relative relationship
between input signals can be observed visually.
(3) Time axis setting
100 μs/DIV to 5 min/DIV
(4) Time axis resolution 100 points/DIV
(5) Memory capacity
(6) Waveform magnification/compression display and print
Time axis direction: ×10 to ×1/10000
Voltage axis direction: ×10 to ×1/2 (single, X-Y single screen)
×5to×1/4 (2 to 16, X-Y4 screen)
With the variable function, vernier function, zoom function
(7) Display format
Time axis waveform: single, dual, quad screen display (LCD), octhex screen
displayprinter only
X-Y waveform: dot, line, singlequad screen
(8) Printing
Auto print, Manual print, Partial print, Report (B4 size) print, Screen hard
copy.
Multiple printing possible.
(9) High-quality print
Smooth print function approximates analog waveform.
(10) Logging function
Numeric printout of waveform data
(11) Memory segmentation function
Helps to reduce dead time. (Sequential save function)
Memory is divided into blocks which can freely selected by the user for storing
measurement data. (Multi-block function)
(12) Processing function
Waveform processing (arithmetic processing, differential processing etc.)
Waveform parameter processing (frequency measurement, rms measurement
etc.)