2.10 Notes on Measurement
────────────────────────────────────────────────────
WARNIN
G
To avoid electric shock accident, before removing or replacing an input
module, confirm that the instrument is turned off and that the power
cord, connection cords, and thermocouples are disconnected.
To avoid the danger of electric shock, never operate the product with an
input module removed. To use the product after removing an input
module, install a blank panel over the opening of the removed module.
Before using the product, make sure that the insulation on the
connection cords is undamaged and that no bare conductors are
improperly exposed. Using the product under such conditions could
result in electrocution. Replace the connection cords specified by Hioki.
CAUTIO
N
・When making measurements on an AC power line for example, using a voltage
transformer, be sure to connect the voltage transformer ground terminal to
ground.
NOTE
・ If the 8937 VOLTAGE/TEMP UNIT’s input terminal is exposed to a strong
draft, loss of thermal equilibrium at the input may result in measurement
error. When taking measurements under such conditions, arrange the unit in
such a manner that the input terminal is protected for direct exposure to
drafts.
・ If ambient temperature changes suddenly, loss of thermal equilibrium can
result in measurement error. When this occurs, allow the unit to acclimate to
the new temperature for about one hour, then take measurements after
thermal equilibrium is reached.