User Manual

500
────────────────────────────────────────────────────
21.3 Reference
────────────────────────────────────────────────────
2
1.3.4 Recorder Function
1234・・・・・・
・・・・・・・10
0
1 DIV
Maximum voltag
e
Minimum voltag
e
One piece of sample dat
a
Data obtained in th
e
set sampling period
1234・・・・・・
・・・・・・・10
0
1 DIV
Maximum voltag
e
Minimum voltag
e
One piece of sample dat
a
Set a short sampling period
One division is equal to 100 samples.
One piece of sample data collected using the recorder function contains the
maximum and minimum voltage obtained in the set sampling period.
Therefore, this data is of a certain width.
When a short sampling period is set and the input waveform changes slightly,
a sudden disturbance such as noise will increase the difference the between
the maximum and minimum values. To eliminate this phenomenon, set a
long sampling period.