21.1 Error Messages
────────────────────────────────────────────────────
W
ARNING 358Change the time axis
range (1 ms)
The anti-aliasing filter setting is invalid for the time
axis range of 100 to 500 μs. Please change the time
axis to between 1 ms and 5 min.
W
ARNING 359Change the frequency
range (40 kHz)
The anti-aliasing filter setting is invalid for the
frequency range of 400 to 80 kHz. Please change the
frequency range to between 40 kHz and 133 mHz.
W
ARNING 380No data in RefBlock. When using the memory segmentation function (multi-
block), there is no data in the reference block.
W
ARNING 381Refblock = Using block When using the memory segmentation function (multi-
block), the reference block and the block specified by th
e
"using block" item are the same.
W
ARNING 382No waveform data Because there is no waveform data present, it cannot b
e
displayed. Start measurement operation to capture
data.
W
ARNING 384Different Refshot The recording lengths are different for the reference
block and the block specified by the "using block" item.
Capture data with the recording lengths set the same.
W
ARNING 386InvalidRECORDER In the recorder function, the key pressed is invalid.
W
ARNING 387InvalidX-Ycont In the X-Y recorder function, the key pressed is invalid
.
W
ARNING 388No comparison AREA No waveform evaluation area. Create waveform
evaluation area.
W
ARNING 389Cannot use, Printer The printer cannot be used when the time axis range is
20 to 200 ms/DIV and CONT.
W
ARNING 390Cannot set over up level Setting cannot be higher than upper limit.
W
ARNING 391Cannot set under low
level
Setting cannot be lower than lower limit.
W
ARNING 395Variable is ON. Since the variable is set to ON, the waveform will
maintain its current size even if the voltage-axis range
is changed.
W
ARNING 396Out of rangevariable The settable range for the variable function (captured
measurement range value×±10000) was exceeded.
When this warning appears, the upper and lower value
setting is automatically changed to be within range.
W
ARNING 397Out of rangescaling POINT was set for scaling and the settable range was
exceeded. See Section 21.3.7.
W
ARNING 398A/B cursor positions
invalid.
Cursor A and B are not located correctly.
W
ARNING 421Equation contains a
syntax error.
Equation contains a syntax error. Correct equation.
W
ARNING 422Cannot copy the equation
.
(Zxx)
Copy function cannot be carried out, because the copy
source equation contains a Z number higher than the
copy target.
W
ARNING 423Upper value has to be
bigger than lower value.
Upper value has to be bigger than lower value.
W
ARNING 425Length of data in memory
is too short.
The recording length is not sufficient for the number of
FFT points.
W
ARNING 610No interface card. Insert the interface card (PC card).
W
ARNING 620Cannot connect. (LAN) Check the settings or connection environment.
W
ARNING 622LAN: Bad Ip address. The IP address is not set correctly.
W
ARNING 623LAN: Bad server Ip
address.
The IP of the server (that runs the 9333) is not specifie
d
correctly.
W
ARNING 624LAN: Can not connect to
server.
The 8826 cannot connect to the PC that runs the 9333.
Check the connection environment.
W
ARNING 625LAN: Can not connect to
9333.
The 8826 cannot connect to the 9333. Check on the PC
to see whether the 9333 is in standby status