21.1 Error Messages
────────────────────────────────────────────────────
W
ARNING 358:Change the time axis
range (1 ms∼)
・The anti-aliasing filter setting is invalid for the time
axis range of 100 to 500 μs. Please change the time
axis to between 1 ms and 5 min.
W
ARNING 359:Change the frequency
range (40 kHz∼)
・The anti-aliasing filter setting is invalid for the
frequency range of 400 to 80 kHz. Please change the
frequency range to between 40 kHz and 133 mHz.
W
ARNING 380:No data in Ref.Block. ・When using the memory segmentation function (multi-
block), there is no data in the reference block.
W
ARNING 381:Ref.block = Using block ・When using the memory segmentation function (multi-
block), the reference block and the block specified by th
e
"using block" item are the same.
W
ARNING 382:No waveform data. ・Because there is no waveform data present, it cannot b
e
displayed. Start measurement operation to capture
data.
W
ARNING 384:Different Ref.shot. ・The recording lengths are different for the reference
block and the block specified by the "using block" item.
Capture data with the recording lengths set the same.
W
ARNING 386:Invalid.(RECORDER) ・In the recorder function, the key pressed is invalid.
W
ARNING 387:Invalid.(X-Ycont) ・In the X-Y recorder function, the key pressed is invalid
.
W
ARNING 388:No comparison AREA. ・No waveform evaluation area. Create waveform
evaluation area.
W
ARNING 389:Cannot use, Printer. ・The printer cannot be used when the time axis range is
20 to 200 ms/DIV and CONT.
W
ARNING 390:Cannot set over up level. ・Setting cannot be higher than upper limit.
W
ARNING 391:Cannot set under low
level.
・Setting cannot be lower than lower limit.
W
ARNING 395:Variable is ON. ・Since the variable is set to ON, the waveform will
maintain its current size even if the voltage-axis range
is changed.
W
ARNING 396:Out of range.(variable) ・The settable range for the variable function (captured
measurement range value×±10000) was exceeded.
When this warning appears, the upper and lower value
setting is automatically changed to be within range.
W
ARNING 397:Out of range.(scaling) ・POINT was set for scaling and the settable range was
exceeded. See Section 21.3.7.
W
ARNING 398:A/B cursor positions
invalid.
・Cursor A and B are not located correctly.
W
ARNING 421:Equation contains a
syntax error.
・Equation contains a syntax error. Correct equation.
W
ARNING 422:Cannot copy the equation
.
(Zxx)
・Copy function cannot be carried out, because the copy
source equation contains a Z number higher than the
copy target.
W
ARNING 423:Upper value has to be
bigger than lower value.
・Upper value has to be bigger than lower value.
W
ARNING 425:Length of data in memory
is too short.
・The recording length is not sufficient for the number of
FFT points.
W
ARNING 610:No interface card. ・Insert the interface card (PC card).
W
ARNING 620:Cannot connect. (LAN) ・Check the settings or connection environment.
W
ARNING 622:LAN: Bad Ip address. ・The IP address is not set correctly.
W
ARNING 623:LAN: Bad server Ip
address.
・The IP of the server (that runs the 9333) is not specifie
d
correctly.
W
ARNING 624:LAN: Can not connect to
server.
・The 8826 cannot connect to the PC that runs the 9333.
Check the connection environment.
W
ARNING 625:LAN: Can not connect to
9333.
・The 8826 cannot connect to the 9333. Check on the PC
to see whether the 9333 is in standby status