21.1 Error Messages
────────────────────────────────────────────────────
1
2
3
4
5
2
1
7
8
9
1
0
1
1
1
2
1
3
1
4
A
W
ARNING 216MO drive selected in auto
-
save.
MO drive selected in auto-save, therefore SCSI cannot
be selected.
W
ARNING 217SCSI selected in auto-
save.
Connected SCSI devise selected in auto-save, therefore
MO disk cannot be selected.
W
ARNING 300Cannot START.
SYSTEM
Cannot start measurement from SYSTEM screen.
W
ARNING 301 InvalidSYSTEM The key pressed is not valid on the system screen.
W
ARNING 324Ignore in running
AVERAGE
Because averaging is used, waveform processing is not
carried out during the start operation.
W
ARNING 325Invalid
(Waveform processing)
The vernier change is disabled for the processed
waveforms.
W
ARNING 327Invalid
COMPARISON
Pressed key is invalid, when waveform evaluation is
being carried out.
W
ARNING 328Invalid
OVER LAY
Operation is not possible, since the overlay function is
enabled.
W
ARNING 329Wrong format for
Comparison
Since the format is not SINGLE or XY single, a
waveform decision is not possible.
W
ARNING 330Cannot set
SHOT too long
The recording length is too long for the memory
segmentation function or a waveform processing
calculation to be carried out.
W
ARNING 334Cannot set.(AVERAGE) Averaging and waveform decision cannot be carried out
.
W
ARNING 335Cannot set
SEQUENTIAL
Waveform processing cannot be carried out, because
memory segmentation function is active.
W
ARNING 336Cannot set
MULTI BLOCK
Waveform processing cannot be carried out, because
memory segmentation function is active.
W
ARNING 337Cannot set
ROLL MODE
Overlay and waveform decision cannot be carried out,
because roll mode is active.
W
ARNING 338Cannot setOVER LAY
Waveform processing cannot be carried out, because
overlay function is active.
W
ARNING 339InvalidSTATUS On the status screen, the key pressed is invalid.
W
ARNING 340Invalid
EXT sampling
External sampling cannot be carried out.
W
ARNING 345Cannot set.
(AND logic trigger)
AND cannot be set between trigger source RMS level
trigger and the logic trigger.
W
ARNING 346Cannot set.
(AND rms level)
AND cannot be set between trigger source RMS level
trigger and the logic trigger.
W
ARNING 347Invalid(Pre Trigger) If no trigger has been set, pre-trigger (0%) is invalid.
W
ARNING 351Cannot setFree run The pre-trigger setting cannot be made, since all trigge
r
sources are switched off (free run).
W
ARNING 352InvalidCHAN On the CHANNEL screen, the key pressed is invalid.
W
ARNING 353Cannot set.
(time/div:100ms - 5min)
The voltage drop trigger cannot be used when the time
axis range is 100ms to 5 min/DIV.
W
ARNING 354Cannot set.
(frequency:133mHz-400Hz)
The voltage drop trigger is available is a frequency-axis
range of 800 Hz to 400 kHz.
W
ARNING 355Change time axis.
(RECORDER)
Since the memory time axis range is delayed, the
recorder time axis is delayed.
W
ARNING 356Change time axis.
(MEMORY)
If the speed of the recorder time axis range is increased
beyond two steps below the memory time axis range,
the memory time axis range speed is incremented.
W
ARNING 357AAF is invalid The anti-aliasing filter setting is invalid for the time
axis range of 100 to 500 s and the frequency range of
400 to 80 kHz.