19.2 Analog Inputs
────────────────────────────────────────────────────
Frequency ranges 0.05, 0.1, 0.5, 1,5, 10, 50, 100, 500 Hz/DIV 1, 5 kHz/DIV
5, 10, 50, 100, 500 r/min/DIV
Power source frequency ranges: 50 Hz (40 to 60 Hz), 60 Hz (50 to 70 Hz)
Frequency accuracy ±0.2%f.s.(except 100 kHzf.s. range)
±0.7%f.s.(100 kHzf.s. range)
±0.032Hz(Power source frequency ranges)
Frequency hold ON/OFF (waiting time 10 ms, 1 s variable )
When hold is OFF, the current measurement value is halved if the next
measurement value is not fixed within the waiting time.
Count ranges 5, 10, 50, 100, 500 counts/DIV
1, 5, 10, 50, 100, 500 k counts/DIV
Duty range 100%f.s.
Duty accuracy ±1%(10 Hz to 10 kHz)
Threshold value (common) -10 to +10 V variable(0.2 V steps)
Frequency measurement
ranges (common)
DC to 100 kHz (Frequency)
DC to 90 kHz (Count)
10 Hz to 100 kHz (Duty)
Response time (common) 10μs not greater (Frequency (more than 300 Hz), Count)
50μs not greater (Frequency (300 Hz not greater), Duty)
The sampling period of the MEMORY HiCORDER in which the unit is
installed must be added to the above.
Voltage range 500μV/DIV, 1, 2, 5, 10, 20, 50, 100, 200, 500 m/DIV 1, 2 V/DIV
Current range □ Using the 9270, 9272 (20A), 9277, 3273, 3273-50
5, 10, 20, 50, 100, 200, 500 mA/DIV, 1 A/DIV
□ Using the 9271, 9272 (200A), 9278
50, 100, 200, 500 mA/DIV, 1, 2, 5, 10 A/DIV
□ Using the 9279
100
*
, 200
*
, 500 mA/DIV
1
*
,2
*
,5,10
*
,20
*
, 50, 100
*
A/DIV
*
:Vertical resolution 64 LSB/DIV
Frequency characteristic
(common)
DC to 400 kHz ±3 dB (DC coupling)
(When using a sensor, depends on the characteristics of the sensor.)
DC amplitude accuracy *
1
(common)
±0.4%f.s. (Using the 9279 ±0.5%f.s.)
Zero position accuracy *
1
(common)
±0.15%f.s. (Using the 9279 ±0.2%f.s.)
Temperature characteristic*
1
(common)
Gain: ±0.025%f.s./℃
Zero position: ±0.03%f.s./℃ (range: more than f.s.=100 mV range)
±0.04%f.s./℃ (range: more than f.s.=50 mV range)
Common mode rejection ratio 80 dB minimum (at 50/60 Hz and with signal source resistance 100 Ω
maximum)
Noise*
1
150μVp-p max (at 0.5 mV/DIV range)
Frequency・Count・Duty Measurement
Voltage・Current Measurement
(
*1
): When measuring current, the accuracy and characteristics of the sensor or probe must be
added to the above.