User Manual

474
────────────────────────────────────────────────────
19.2 Analog Inputs
────────────────────────────────────────────────────
Es Signal voltage
Rs Signal source resistanc
e
Rin Input resistance
Rin
Rs+Rin
V
Measurement errorEs 1-
R
s
Ri
n
E
s
1
9.2 Analog Inputs
Measurement Errors Caused by Signal Source Internal Resistance
If the signal source impedance is higher than the input impedance of the
unit, a measurement error will occur.
The input impedance of the 8936 ANALOG UNIT is 1 MΩ. If the signal
source impedance is 1 kΩ, an error of about 0.1% will occur.