User Manual
23
────────────────────────────────────────────────────
2.6 9018-10, 9132-10 CLAMP ON PROBE Connection
────────────────────────────────────────────────────
DANGE
R
Clamp-on probe should only be connected to the secondary side of a
breaker, so the breaker can prevent an accident if a short circuit occurs
.
Connections should never be made to the primary side of a breaker,
because unrestricted current flow could cause a serious accident if a
short circuit occurs.
Before using the 9018-10, 9132-10 CLAMP ON PROBE, be sure to
carefully read its instruction manual and familiarize yourself with the
operation principles of this product.
Connector ridge
BNC connector counterclockwise
NOTE
DANGE
R
When using grabber clips, the 9322’s maximum rated to-ground voltage
is 1,500 V AC or DC; when using alligator clips, it is 1,000 V AC or DC.
To avoid electrical shock and possible damage to the unit, never apply
voltages greater than these limits between the input channel terminals
and chassis, or across the inputs of two 9322s.
The maximum input voltage of the 9322 is 2000 V DC, 1000 V AC.
Attempting to measure voltage in excess of the maximum input could
destroy the product and result in personal injury or death.
2
.6 9018-10, 9132-10 CLAMP ON PROBE Connection
2
.7 9322 DIFFERENTIAL PROBE Connection
The HIOKI 9018-10, 9132-10 CLAMP ON
PROBE can be used to perform current
measurement. This probe has a voltage
output. The 8826 is designed for voltage
input. It cannot be used to perform current
measurements by itself.
Connections
Connect the BNC connector of the 9018-10,
9132-10 CLAMP ON PROBE to the analog
input of the 8826, as shown in the
illustration.
・ The 8826 will indicate measurement results as voltage.
・ Using the scaling function, units can be converted to "A" for display.
See Section 9.7.2.
・ When the clamp-on probe is used for measurement, the measurement
precision will be affected both by the 8826 precision and clamp-on probe
precision ratings. The same is true for cases where other clamps are used.
・ When using the 9018, 9132 CLAMP ON PROBE, always use the 9199
CONVERSION ADAPTOR.
For details, see the instruction manual of the 9322 DIFFERENTIAL PROBE.