User Manual

462
────────────────────────────────────────────────────
18.7 Recorder & Memory Function
────────────────────────────────────────────────────
Time axis Recorder
20, 50, 100, 200 ms/DIV (display only)
500 ms/DIV
1, 2, 5, 10, 30 s/DIV
1, 2, 5, 10, 30 min/DIV
1 h/DIV
Memory Recorder
100, 200, 500 μs/DIV
1, 2, 5, 10, 20, 50, 100, 200, 500 ms/DIV
1, 2, 5, 10, 30 s/DIV
1, 2, 5 min/DIV
Time axis resolution 100 points/DIV
Sampling period 1/100 of the time axis.......Memory waveform
Recording length Standard (16 M):
25, 50, 100, 200, 500, 1000, CONT (Recorder waveform)
25, 50, 100, 200, 500, 1000, 2000, CONT (Memory waveform)
Expansion (64 M):
25, 50, 100, 500, 1000, 2000, 5000 DIV, CONT (Recorder waveform)
25, 50, 100, 500, 1000, 2000, 5000, 10000 DIV (Memory waveform)
Screenprint format The styles single, dual, quad,DISPLAY screen, octhexPrint only,
Display quad style
Recording line display 12-colorLCD, Printout: 4-type
Display Switchable between recorder and memory waveforms
Printer output During measurement operation, recorder waveform only. After data
capture, printout of recorder waveform as on display or memory
waveform.
Waveform storage (REC) Last 1000 divisions of data saved in memory
(*1)
Can be checked by reverse scrolling and reprinted
Additional recording function ON/OFF*
2
Trigger source CH1 to CH32, CHA to CHH and external trigger (Memory waveform)
Report print B4 size print available
Variable function Provided
Zoom function Provided( in the memory recoder function )
Real time save function Time axis range: 100 ms/DIV( less than 16 ch)
200 ms/DIV to 5 min/DIV (more than 17 ch)
(1 DIV = 100 points)
Save destination: PC card or 9598 MO Drive Unit (option)
Save time: depends on available space on the PC card or MO disk
1
8.7 Recorder & Memory Function
*1
Expanded to 64 words: 5000 DIV
*2
Additional recording function (recording data without paper)
When enabled, the memory is regarded as printer paper. Recording starts at the end of previous
data, without erasing them. When the 2000 DIV
*1
has been reached, old data will be overwritten.
When OFF, previous data will be erased. Set to ON if erasing is not desired.