User Manual
461
────────────────────────────────────────────────────
18.6 RMS Recorder Function
────────────────────────────────────────────────────
Time axis 5, 10, 30 s/DIV
1, 2, 5, 10, 30 min/DIV
1 h/DIV
Time axis resolution 100 points/DIV (with the printer)
Number of channel Analog 32 ch + logic 32 ch
Sampling period 20 rms data/s (200 μs fixed)
*1
RMS accuracy ±3% f.s. (at 50/60 Hz±2 Hz, DC) (f.s.=20 DIV)
Measuring object Commercial power supplies (50/60 Hz), DC
Recording length ・Standard (16 M): 25,50,100,200,500,1000,2000 DIV,CONT
*2
・Expansion (64 M): 25,50,100,200,500,1000,2000,5000,10000 DIV,CONT
*
2
・Arbitrarily (Set from 1 division to the maximum number of divisions a
t
1-division intervals.)
Screen・print format The styles single, dual, quad,(DISPLAY screen), oct・hex(Print only,
Display quad style)
Interpolation Line
Recording line display 12-color(LCD), Printout: 4-type
Waveform magnification/
compression
・Time axis
×1, ×1/2, ×1/5, ×1/10, ×1/20, ×1/50, ×1/100, ×1/200, ×1/500
・Measurement range
×10, ×5, ×2, ×1, ×1/2 (single)
×5, ×2.5, ×1, ×1/2, ×1/4 (2, 4, 8, 16)
Waveform storage ・Last 2000 divisions of data saved in memory
*2
・Can be checked by reverse scrolling and reprinted
Print function ON/OFF and reprinted
Report print B4 size print available
Additional recording function ON/OFF
*3
Variable function Provided
Logging function Records measured data as digital values
1
8.6 RMS Recorder Function
(
*1
):Refer to 21.3.5.
(
*2
):Expanded to 64 words: 10000 DIV
(
*3
):Additional recording function (recording data without paper)
When enabled, the memory is regarded as printer paper. Recording starts at the end of previous
data, without erasing them. When the 2000 DIV
*2
has been reached, old data will be overwritten.
When OFF, previous data will be erased. Set to ON if erasing is not desired.