17.6 GO/NG Evaluation Output Terminal
────────────────────────────────────────────────────
16
17
18
19
5
6
7
8
9
1
0
1
1
1
2
1
3
1
4
A
Output Terminal
Evaluation result
GO NG
GO
Low level High level
NG
High level Low level
10 kΩ
5V
GO, NG
4.7 kΩ
4.7 kΩ
HIGH
4.0 to 5.
0V
LOW
0to0.
5V
min. 70 ms
GND
Evaluation output interval
GO, NG
GND
Reset
1μF
5V
D
R
1to5kΩ
5to24V
Output
Relay
2SC1815 etc.
HC74
CK
Q
HC04
PR
10 kΩ
1
7.6 GO/NG Evaluation Output Terminal
When waveform evaluation or waveform parameter evaluation is used, a
signal is output from these connectors when the result is GO (pass) or NG
(fail).
Output signal
Signal type Open-collector signal, active Low
Output voltage range High level: 4.0 to 5.0 V, Low level: 0 to 0.5 V
Maximum input voltage -20 to +30 V, max. 500 mA, max. 200 mW
Evaluation output interval (min. 70 ms)
The evaluation outputs are shown in the following table. Between these states,
there is an interval during which the next data are read and waveform data
are created. The duration of this interval is inversely proportional to the time
axis and proportional to the recording length.
The following diagram shows an example of a circuit that operates an alarm