User Manual

448
────────────────────────────────────────────────────
17.3 External Trigger Output Terminal [TRIG OUT]
────────────────────────────────────────────────────
10 kΩ
5V
TRIG.OUT
4.7 kΩ
4.7 kΩ
HIGH
4.0 to 5.
0V
LOW
0to0.
5V
min. 2 ms
GND
NOTE
1
7.3 External Trigger Output Terminal [TRIG OUT]
When triggering occurs, a signal is output from this terminal.
Several 8826 units can be synchronized for parallel operation.
Trigger output signal
Signal type Open-collector signal, active Low
Output voltage High level: 4.0 to 5.0 V
range Low level: 0 to 0.5 V
Pulse width Low level:min. 2 ms.
Maximum input voltage -20 to +30 V, max. 500 mA, max. 200 mW
When the auto range function is activated by pressing the AUTO key, a trigger
output signal is generated. This should be taken into consideration when
using both the trigger output and the auto range function. (Memory recorder
function only)