User Manual

446
────────────────────────────────────────────────────
17.1 Connecting the Terminals
────────────────────────────────────────────────────
φ1.0 mm
(can be used 0.3 to 1.0 mm dia.)
10 mm
10 mm
0.75 mm
2
(
can be used 0.3 to 0.75 mm
2
D
iameter per strand:
minimum 0.18 mm dia.
Single strand
Multi-strand
Recommended wire Single strand: 1.0 mm dia. (AWG #18)
Multi-strand: 0.75 mm
2
Usable limits Single strand: 0.3 to 1.0 mm dia. (AWG #26 to #18)
Multi-strand: 0.3 to 0.75 mm
2
(AWG #22 to #20)
Strand diameter: minimum 0.18 mm
Standard insulation stripping length: 10 mm