16.3 Waveform GO/NG Evaluation
────────────────────────────────────────────────────
1
6.3.3 Setting the GO/NG Stop Mode
Screen: STATUS1 (MEM
)
: STATUS2 (FFT)
Move the flashing cursor to the position
shown in the figure on the left.
Make the setting with the function keys.
:Stop recording on GO result.
Selection
Selection
Method
Method
:Stop recording on NG result.
:Stop recording on GO or NG result.
NOTE
When waveform evaluation is enabled (OUT or ALL OUT is selected), the
"Stop Conditions" menu appears. Specify which evaluation option, GO or NG,
should be used to stop the recording.
When waveform parameter measurement and waveform evaluation are carried
out simultaneously, the screen displayed waveform evaluation results. Check
the results of parameter evaluation using "
*
" (NG decision) accompanying the
figure. Alternatively, a beeping sound can be used for the evaluation instead
of the "
*
" mark; the machine beeps if the result of either parameter or
waveform evaluation is NG (MEM).