429
────────────────────────────────────────────────────
16.2 Calculating Waveform Data
────────────────────────────────────────────────────
・Press the FILE key to call up the FILE screen.
For details on LOAD, refer to 14.8.
Internal memory data
・This setting calculates the measurement data between A and B
cursors.
When there is no A and B cursors, calculates all data.
For details on A/B cursors, refer to Section 11.2.
・Printout can be carried out.
・Saving can be carried out.
・Press the STATUS key to call up the STATUS4 screen.
For details on setting.
Displaying calculation result
・Displays waveform calculation result on the screen.
Select the calculating data
Setting the items
Operating the result
Measurement end
Loading from media
・Waveform calculation execute the stored measurement data.
Setting the calculating range
・Move the flashing cursor to wave calculation. And use the
function key to select "exec".
・Displays "Waveform calculating".
Waveform calculation
NOTE
Waveform processing of data in internal memory or media
・The maximum recording length allowing waveform processing calculation is
1000 divisions (5000 divisions for 64 M words).
・When using the trace cursor, the trace point value is displayed as processed
value.
・When the cursors overlap, processing is carried out for that point.
・Line cursor can not be set calculating ranges.
・When only the cursor A is used, the waveform data from the position of
cursor A to the end of the data is calculated.
・The result can be recalculated by altering the calculation process.
・A stored waveform read by the Recorder & Memory function can be
subjected to waveform processing by the Memory function, but doing so
erases the Recorder waveform.