16.2 Calculating Waveform Data
────────────────────────────────────────────────────
1
6.2.1 Preparing for Waveform Processing
Screen: STATUS4 (MEM
)
Press the STATUS key to call up the
STATUS4 screen.
Move the flashing cursor to the position
shown in the figure on the left.
Make the setting with the function keys.
:Disable waveform processing.
:Enable waveform processing.
The various setting items are
displayed.
Selection
Selection
Method
Method
:Execute Waveform Processing.
NOTE
・The maximum recording length at which waveform processing calculation is
possible is 1000 divisions (5000 divisions for 64 M words).
If the recording length exceeds this limit, the waveform processing
calculation is disable.
・When the memory segmentation function or roll mode is used, waveform
processing is not possible.
・The averaged waveforms can be processed following the measurement.
・When scaling is set for the channel in which the processing result is to be
stored, scaling is not carried out and only the unit is valid. For details on
scaling, refer to Section 9.7.
・When the waveform processing calculation executes simultaneously with
data acquisition, a forced termination displays the results being calculated.
In such cases, press the function key (
exec) to reexecute the calculation.