16.1 Waveform Parameter Calculation
────────────────────────────────────────────────────
Press the FILE key to call up the FILE screen.
For details on LOAD, refer to Section 14.8.
Internal memory data
This setting calculates the measurement data between A and B
cursors.
When there is no A and B cursors, calculates all data.
For details on A/B cursors, refer to Section 11.2.
Result to printer ON →Prints the evaluation results.
Result save
ON
 →
Saves the evaluation results.
Press the STATUS key to call up the STATUS3 screen.
For details on setting, refer to Section 16.1.1.
Depending on the results of the waveform parameter
calculation, a GO (pass) or NG (fail) result is returned.
When the evaluation result is GO
Outputs GO signal.
When the evaluation result is NG
Outputs NG signal.
the calculation value for that channel is marked with an "
*
"
(on the display and the printout).
When COMP setting: ON
Displaying calculation result
Displays parameter calculation result on the screen.
Select the calculating data
Setting the items
Parameter evaluation
Operating the evaluation resul
t
Measurement end
Loading from media
Parameter calculation execute the stored measurement data.
Setting the calculating range
Move the flashing cursor to measurement. And use the
function key to select "exec".
Displays "Calculating".
Parameter calculation
NOTE
Parameter calculation of measurement data loaded from media
Line cursor (horizonal) can not be set calculating ranges.
When only the cursor A is used, the waveform data from the position of
cursor A to the end of the data is calculated.
The result can be recalculated by altering the calculation process.