16.1 Waveform Parameter Calculation
────────────────────────────────────────────────────
16
17
18
19
5
6
7
8
9
1
0
1
1
1
2
1
3
1
4
A
NOTE
1
6.1.3 Setting Waveform Parameter Evaluation GO/NG Stop Mod
e
Screen: STATUS
3
Move the flashing cursor to the position
shown in the figure on the left.
Use the function keys or the JOG control
to make the selection.
Method
Method
Selection
Selection
Stop recording with NG result.
Stop recording with either GO or
NG result.
Stop recording with GO result.
1
6.1.4 Executing Waveform Parameter Calculation
When waveform parameter measurement and waveform evaluation are
carried out simultaneously, the screen displayed waveform evaluation
results. Check the results of parameter evaluation using "
*
" (NG decision)
accompanying the figure. Alternatively, a beeping sound can be used for the
evaluation instead of the "
*
" mark; the machine beeps if the result of either
parameter or waveform evaluation is NG.
When the evaluation result is NG, the calculation value for that channel is
marked with an "
*
" (on the display and the printout).
The result of the evaluation is NG if any of the values is NG.
While all evaluation results appear on the display screen, the evaluation
results are printed out for each parameters in the print mode.
When the evaluation result is NG, an NG output signal can be obtained
between the NG terminal and the GND terminal. For details, please refer
to the Section 17.6.
When waveform parameter evaluation is ON, CH.SET key settings are
automatically OFF.
When ALL is selected using the A and B cursors, the calculation No.4 result
will not be displayed on the screen.
Set which evaluation result (GO or NG) shall be used to stop recording.
Calculation is carried out in the order No.1 through No.4.
Also for channels where no input unit is installed, parameter calculation is
carried out if waveform processing results or data loaded from media are
stored in the channel.
The scaling setting has effect. (RMS value and area value are calculated
after scaling.)
When measurement is set to ON, waveform data after waveform processing are
used for parameter calculation.