15.2 Using the Multi-Block Function
────────────────────────────────────────────────────
NOTE
Screen: STATUS
2
Press the STATUS key to call up the
STATUS2 screen.
Move the flashing cursor to the position
shown in the figure on the left.
Press the F3 [ MULTI ].
Disable memory segmentation.
Enable sequential save function.
Selection
Selection
Method
Method
Enable multi-block function.
(1) Select the Memory Segmentation
Move the flashing cursor to the division
item, as shown in the figure on the left.
Use the JOG control or the function keys
to make the selection.
Move the cursor up in the selection
window.
Move the cursor down in the
selection window.
Selection
Selection
(2) Set the number of memory block
While the multi-block function is being used, the waveform processing
calculation and averaging functions are disabled.
If the blocks have different recording lengths, the overlap block display is
disabled.
The VIEW key can be used to change the displayed memory block or to call
up information about the usage status of memory blocks. See Section 11.6.