15.1 Using the Sequential Save Function (MEM)
────────────────────────────────────────────────────
Press the START key and LED light.
Data recording starts when trigger condition are met.
start block
Start block is stored.
wave display O N: start block waveform is displayed.
wave display OFF: waveform is displayed.
end block
Displays end block waveform.
When auto saving is enabled, waveforms in all blocks are saved in
a batch.
Auto print
Auto save
End block is stored.
wave display O N: end block waveform is displayed.
wave display OFF: waveform is displayed.
When auto printing is enabled, all blocks are printed from the start
block while individual blocks are displayed.
(MEM) Each time when trigger conditions are met, data are recorded
and memory contents are overwritten.
E
nd of measurement
I
f the STOP key is pressed during recording, recording continues
u
ntil the end of the current block.
(
Waveform display, auto printout and auto save are executed.)
When the STOP key is pressed twice during measurement, the
8826 is forcibly stopped. (Auto printout and auto save are not
executed.)
(
REC&MEM) Measurement is started again. The memory waveform
data is cleared.
E
nd of measurement
W
hen the STOP key is pressed once during measurement, the 882
6
is forcibly stopped.
Measurement Start
Measurement Start
Measurement end
Measurement end
Recording end
Recording end
Trigger mode:SINGLE
Trigger mode: REPEAT
AUTO
When recording with a preset time interval, recording continues until
t
he preset stop time.
I
f the STOP key is pressed during measurement, measurement is
t
erminated.
The memory waveform is cleared after each preset time period.
Trigger mode: TIMER
(REC&MEM only)
End of measurement
(
MEM) When the STOP key is pressed twice during measurement,
the 8826 is forcibly stopped. (Auto printout and auto save are not
executed.)
(
REC&MEM) When the STOP key is pressed once during
measurement, the 8826 is forcibly stopped.
When the STOP key is pressed twice during measurement, the
8826 is forcibly stopped. (Auto printout and auto save are not
executed.)
NOTE
Relation between trigger mode and sequential save function
When the following waveform display (memory recorder) is on and the view
function is on, the dead time becomes very large