14.8 LOAD
────────────────────────────────────────────────────
NOTE
・In case of "OVERWRITE," the settings in the unit are given preference.
When the data in the unit differs from the file functions or time-axis range,
the data cannot be loaded.
・"OVERWRITE" is not possible with data measured using external sampling.
・The data loading turn: CH1→32, CHA→H
・If the plural data are loaded in the same channel, the data most recently
sent remains in the memory.
・When batch saving with the Recorder & Memory function, the measurement
data (with REC, MEM extensions) are created together with an index file
(R_M). When only the measurement data are read, these are read to the
respective functions. To read to the Recorder & Memory, read the index file.
・Do not change or delete the name of a file to be batch read (R_M). The batch
reading may not be executed correctly.
・Data stored in the text format is not readable by the 8826.
Auto setup function
When the power is switched on with a floppy disk or a PC card inserted, the
setting up of the 8826 is performed automatically by reading the setting data
file called "STARTUP.SET" and the evaluation area file called "STARTUP.ARE" on
the floppy disk.
Waveform evaluation settings depend on "STARTUP.ARE"
The same measurement conditions can be simply established by using this
function.
■ Method
Define the setup data file name as "STARTUP.SET" and area data file name as
"STARTUP.ARE". Save the file to the root directory of the floppy disk or a PC
card. The file is read to the 8826 when the equipment is turned on.
・If "STARTUP.SET" is present on both the floppy disk and PC cards, the FD
file is used