User Manual

13
────────────────────────────────────────────────────
2.3 Power On/Off
────────────────────────────────────────────────────
1
2
3
4
5
6
7
8
9
1
0
1
1
1
2
1
3
1
4
A
WARNIN
G
Check the following points before the power switch is turned on.
Power supply matches Rated supply voltage (100 to 240 VAC: Voltage
fluctuations of 10% from the rated supply voltage are taken into
account.) and rated supply frequency (50/60 Hz).
The Unit is correctly installed (Section 2.1).
Power cord is correctly connected.
The unit is properly grounded (Section 2.2).
NOTE
2
.3 Power On/Off
After the Power switch is turned on, wait approximately 30 min for the
8936, 8938, 8940 and approximately 1 h for the 8937, 8939, 8947 to stabilize
the inside temperature of the connected input unit in order to obtain
accurate waveforms. Then, make a zero adjustment of the analog unit or
perform an auto balancing of the strain unit prior to measurement.
Zero adjustment See section 9.5.1
Auto balancing See section 9.12
When the unit is turned off, it memorizes the currently used settings and
reestablishes the same settings the next time the unit is turned on again.