13.6 Printing Procedure
────────────────────────────────────────────────────
1
3.6.4 Partial Print
Screen: DISPLAY
Press the DISP key to call up the
DISPLAY screen.
Position cursor A at the start point of the
range to be printed.
Specify the end point of the range with
cursor B. (Move cursor B (or A) to the
right. When the cursor is at the rightmost
edge of the screen, the waveform scrolls to
the left, and cursor A (or B) scrolls with
it.)
Specify the end point of the range with
cursor B. Then press the PRINT key.
The specified range is printed, also if
cursor A is currently off screen.
Method
Method
NOTE
・This function prints the waveform between the A and B cursors (vertical or
trace cursors).
MEM, REC&MEM
Specified range (out of entire data recorded from a measurement) is printed.
REC, RMS
Specified range (out of last 2,000 (expended 64 M is 10,000) divisions
(magnification: ×1) of data in memory) is printed.
REC&MEM
Specified range (out of last 1,000 (expended 64 M is 5,000) divisions is
printed.
・The function is available also when the A/B cursors are currently outside
the range displayed on screen.
・For details regarding the use of the A/B cursors, refer to Section 11.2.
・Printing is possible also when the print format is currently set to "numeric".
・The range is printed also if the positions of cursor A and cursor B are
reversed.
・When only the cursor A is used, the waveform data from the position of
cursor A to the end of the data is printed.
・On the waveform evaluation screen, cursors A and B are available, but
partial printing cannot be performed.
・On the X-Y screen and in FFT function, the partial printing cannot be
performed.