User Manual

355
────────────────────────────────────────────────────
13.6 Printing Procedure
────────────────────────────────────────────────────
1
3.6.1 Manual Print
After the measurement is completed, press
the PRINT key.
Since data are stored, they can be printed
as often as desired.
When magnification/compression was used,
the printout reflects this condition.
If you want to stop the printing, press the
STOP key.
Screen: DISPLAY
Method
Method
NOTE
This mode serves to print waveform data from the internal memory.
MEM, REC&MEM
Measurement data from one measurement (entire recording length) are
printed.
REC, RMS, REC&MEM
The measurement data most recently stored in memory is printed out in up to
2000 divisions (10,000 divisions for 64 M words). In the REC&MEM mode,
data in up to 1000 divisions (5000 divisions for 64 M words) can be printed
out.
FFT
For data recorded in one measurement, all the results of FFT operation are
printed out in their entirety.
When cursors A and B are enabled, partial printing is executed. (excluding
FFT)
When the recording length is set to "CONT" and at least 2000 divisions
(10,000 for 64 M words) are recorded by the recorder or RMS recorder, or at
least 1000 divisions (5000 for 64 M words) are recorded by the REC&MEM
recorder, the time displayed does not represent the value counted from the
trigger point if the time value of "the system 1" is set to "TIME". When it is
set to "DATE," the time at which printing is initiated is printed out.
In waveform evaluation setting (MEM), printing is performed for a single
display screen (25 DIV). To print a waveform with a recording length of
over 25 DIV, compress the time axis. See section 11.3.