13.4 Setting the SYSTEM Screen (printout)
────────────────────────────────────────────────────
1
3.4.2 SCALING Screen
(1) Conversion ratio method
Scaling is performed by specifying a
physical quantity to correspond to a 1V
input signal (conversion ratio: eu/v), an
offset value, and the unit (eu: engineering
units). This will cause the measurement
voltage to be converted into the selected
units.
(2) 2-point method
Scaling is performed by specifying two
input signal points (voltage values) and
the conversion values for these two points
in engineering units (eu). This will cause
the measurement voltage to be converted
into the selected units.
NOTE
1
3.4.3 COMMENT Screen
(1) Title comment input
Title comments of up to 40 characters can
be included on the recording paper.
(2) Input of comments for analog channel
Comments of up to 40 characters can be
included on the recording paper on each
channel.
(3) Input of comments for logic channel
Comments of up to 40 characters can be
included on the recording paper on each
channel.
NOTE
・The scaling function can be used to convert an output voltage from a sensor
or similar into a physical quantity.
・Two types of scaling functions are available.
・The gauge scale (maximum and minimum values of horizontal axis) and A/B
cursor measurement values are displayed in the scaled units.
・Scaling can be performed for every channel.
Refer to Section 9.7.
Three types of comment are available.
If "COMMENT" or "SET & COMMENT" is selected, this comment will be
included on the recording paper in all functions.
Refer to Section 9.8.