User Manual
302
────────────────────────────────────────────────────
11.2 Using the A/B Cursors
────────────────────────────────────────────────────
1
1.2.4 Using the A/B Cursors (X-Y Screen)
Screen: X-Y DISPLAY
Move the flashing cursor to the csr item.
Use the function keys to make selection.
:Disable A/B cursors
:Line cursor (vertical: t, 1/t)
Selection
Selection
Method
Method
:Line cursor (horizontal: v)
:Trace cursor (the point where the
cursor crosses the waveform)
Use the function keys to select the cursor
to be moved.
:Use A cursor only
:Move the A cursor only
Selection
Selection
:Move the B cursor only
:Move both the A and B cursors
The graph setup menu appears below the
cursor. Using the function keys or the
JOG control, move the flashing cursor to
set the graph of the waveform for which
the voltage is to be read. This channel
setting must be performed for both cursors
A and B.
Press the A・B CSR SELECT key so that
the A.B CSR LED lights up.
Rotate the JOG/SHUTTLE control to move
the cursor.
NOTE
The cursor can be used on the X-Y screen. Partial X-Y plotting enables
operation of the A/B cursors.
By specifying a different graph for the A and B cursors, a potential difference
between the waveforms in the respective channels can be determined.