11.2 Using the A/B Cursors
────────────────────────────────────────────────────
1
1.2.3 Using the Trace Cursors
Screen: DISPLAY
Move the flashing cursor to the csr item.
Use the function keys to select the "Trace
cursor".
:Disable A/B cursors
:Line cursor (vertical: t, 1/t)
Selection
Selection
Method
Method
:Line cursor (horizontal: v)
:Trace cursor (the point where the
cursor crosses the waveform)
Use the function keys to select the cursor
to be moved.
:Use A cursor only
:Move the A cursor only
Selection
Selection
:Move the B cursor only
:Move both the A and B cursors
The channel setup menu appears below the
cursor. Using the function keys or the JOG
control, move the flashing cursor to set the
channel of the waveform for which the value
is to be read. This channel setting must be
performed for both cursors A and B.
Using A cursor only
When ALL is selected, the values of all
channels are read.
Press the A・BCSRSELECT key so that the
A.B CSR LED lights up.
Rotate the JOG/SHUTTLE control to move
the cursor.
Used to read the value at point where the cursor crosses the waveform of the
specified channel. A/B cursor can be specified a different channel.