10.12 Manual Trigger
────────────────────────────────────────────────────
Press the TRIG key to call up the
TRIGGER screen.
Move the flashing cursor to the position
shown in the figure.
Use the function keys to make the
selection.
Screen: TRIGGER
Method
Method
:Manual trigger is disabled.
:Manual trigger is enabled.
Selection
Selection
TRIGGER screen (MEM
)
【 Rear panel 】
Trigger Output termina
l
NOTE
1
0.12 Manual Trigger
1
0.13 Trigger Output Terminal
・Triggering occurs when the MANU TRIG key is pressed.
・The manual trigger is always activated when the MANU TRIG key is pressed,
regardless of trigger source AND/OR linking setting.
・When triggering occurs, a signal is output from the TRIG OUT terminal.
・This can be used to synchronize several 8826 units.
・For details, see Section 17.3.
Pressing the AUTO key to use the auto ranging function causes a trigger
signal to be output. Care is therefore required when using the auto ranging
function when the trigger output terminal is in use. (Only a memory recorder
function)