10.11 Timer Trigger Function
────────────────────────────────────────────────────
Move the flashing cursor to the interval
item, as shown in the figure on the left.
Use the function keys or the JOG control
to make the setting.
:Value up, 10-units
:Value up, 1-units
:Value down, 1-units
:Value down, 10-units
Selection
Selection
(3) Set the trigger interval.
E
nd tim
e
R
ecording leng
th
T
ime interv
al
M
EM and FF
T
R
EC and RM
S
■ Note on use for the timer trigger
(1) Set the present time on the system screen first, then set the timer trigger.
(2)SetthestarttimeandendtimetoapointafterthepressingoftheSTART key.
(3) When the trigger mode is set to "SINGLE", only one trigger event is valid
for the start time. The settings for time interval and end time are invalid
(when timer trigger only is ON).
(4) To perform recording at regular intervals, establish the following settings.
Trigger mode: continuous. Other trigger sources: all OFF
Between end of recording and standby, triggers are invalid while
processing is taking place (auto save, auto print, waveform display process,
calculation process) and therefore data may not be recorded at regular
intervals with some measurement setting.
(5) When trigger sources have been linked with OR
All trigger sources are valid. Therefore triggering may occur even before
the start time, and trigger standby and measurement may continue even
after the stop time.
(6) When trigger sources have been linked with AND
・Trigger standby is enabled at the measurement start time and at each
time interval, and triggering will occur when all trigger conditions other
than timer trigger are met simultaneously.
・Set the time interval to 0 can be made active for the preset time interval only.
・When the recording length (recording time) is longer than the time interval:
Timer trigger, which is determined from the time interval during
recording, is disabled.
End time
MEM, FFT:Retrieve measured data equivalent to the recording length
and discontinue measurement.
REC, RMS:Stop retrieving measured data at the set stop time.