10.9 Logic Trigger
────────────────────────────────────────────────────
Move the flashing cursor to the position
shown in the figure on the left.
Select 1 - 4 with the function key display.
Make the setting with the function keys.
When "CH A" is selected, the setting is for
logic input CH A1 - CH A4 from left to
right.
:Signal disregarded
:Low level signal
Selection
Selection
:High level signal
(3) Set the trigger pattern
NOTE
[OR
]
Trigger patter
n
CHA1
1
CHA2 0
CHA3 ×
CHA4 ×
Trigger patter
n
CHA1
1
CHA2 0
CHA3 ×
CHA4 ×
[ AND
]
・If the conditions are met already when measurement is started (AND: all
trigger patterns are met, OR: one trigger pattern is met), triggering does not
occur at this slope. Triggering only occurs if the conditions are removed and
then met again.
・’AND’ cannot be set between the trigger sources using the RMS level trigger
and the logic trigger. When a shift is made from "OR" to "AND," the logic
trigger setting is turned OFF.
Setting example of the Logic Trigger
(1) If the trigger pattern has been set to "10××" with the operator OR, then
triggering occurs as shown in the figure below.
(2) If the trigger pattern has been set to "10××" with the operator AND, then
triggering occurs as shown in the figure below.