User Manual

285
────────────────────────────────────────────────────
10.9 Logic Trigger
────────────────────────────────────────────────────
Use the TRIG key and the CH.SET key to
call up the screen shown in the figure.
Move the flashing cursor to the position
shown in the figure.
Use the function keys to select the setting.
Screen: TRIGGER
Method
Method
Logic trigger is not used.
Triggering occurs if any one of th
e
logic input signals conforms to th
e
trigger pattern.
Triggering only occurs if all of th
e
logic input signals conforms to th
e
trigger pattern.
Selection
Selection
Move the flashing cursor to the position
shown in the figure on the left.
Use the JOG control or the function keys
to make the selection.
In the MEM, REC&MEM, FFT Function
OFF : Trigger filter is disabled
0.1 to 10.0 : Trigger filter is enabled.
Filter width is specified using divisions.
In the REC and RMS Function
Trigger filter is disabled.
Trigger filter is enabled.
Filter width is 10 ms.
Selection
Selection
(1) Set the logic trigger AND/OR linking.
(2) Set the trigger filter
1
0.9 Logic Trigger
The signal of a logic channel can be used as trigger source.
A trigger pattern and logical operator (AND/OR) are specified, and triggering
occurs when the trigger conditions are met.
A trigger filter can be specified, so that triggering occurs only when the trigger
conditions are met within the filter width.