User Manual
278
────────────────────────────────────────────────────
10.8 Analog Trigger
────────────────────────────────────────────────────
Move the flashing cursor to the position
shown in the figure.
Use the function keys or the JOG control
to set the trigger level. The rms value as
well as the trigger level are displayed.
:Value up, large step
:Value up, small step
Selection
Selection
:Value down, small step
:Value down, large step
(3) Set the trigger level.
Setting Item Setting Method
Trigger mode Function key
Trigger type Function key
Channel Function key, Jog (VALUE select key)
Trigger level Function key, Jog (VALUE select key)
RMS value
Pre-trigger Function key, Jog (VALUE select key)
Restriction:The frequency cannot be set.
The selected window is not displayed in the pre-
trigger setting.
1.000 V
1.000 V
1/2 cycl
e
Settings on the DISPLAY screen
Example for Voltage Drop Trigger
For a signal such as shown in the illustration (frequency 50 Hz), the following
applies:
V max. < 1.000 V (RMS < 707.1 mV rms)
To perform trigger measurement under the above conditions, make the
following setting.
Voltage drop f = 50 Hz Level = 1.000 V (707.1 mV rms)