User Manual

275
────────────────────────────────────────────────────
10.8 Analog Trigger
────────────────────────────────────────────────────
1
0.8.2 Window-In, Window-Out Trigger
Upper lim
it
Lower lim
it
Upper lim
it
Lower lim
it
Window-in-trigger
Window-out-trigger
Call up the TRIGGER or DISPLAY screen
.
Move the flashing cursor to the position
shown in the figure.
Use the function keys to select IN or OUT.
Value down, small step
Value down, large step
Move the flashing cursor to the position
shown in the figure on the left, and use
the function keys or the JOG control to se
t
the lower trigger level.
Set the upper trigger level same method.
The upper trigger level must not be
smaller than the lower trigger level, or the
lower trigger level must not be larger than
the upper trigger level.
Value up, large step
Value up, small step
Screen: TRIGGER, DISPLAY
Method
Method
(1) Select the window-in or window-out trigger.
(2) Set the lower and upper trigger levels.
Window-In Trigger
Set upper limit level and lower limit level and activated when the input signal
enters the range between these limits.
Window-Out Trigger
Set upper limit level and lower limit level and activated when the input signal
leaves this range.