User Manual

274
────────────────────────────────────────────────────
10.8 Analog Trigger
────────────────────────────────────────────────────
Move the flashing cursor to the position
shown in the figure on the left.
Use the JOG control or the function keys
to make the selection.
In the MEM, REC&MEM, FFT Function
OFF : Trigger filter is disabled
0.1 to 10.0 : Trigger filter is enabled.
Filter width is specified using divisions.
In the Recorder Function
Trigger filter is disabled.
Trigger filter is enabled.
Filter width is 10 ms.
Selection
Selection
(4) Set the trigger filter
Setting Item Setting Method
Trigger mode Function key
Trigger type Function key
Channel Function key, Jog (VALUE select key)
Trigger level Function key, Jog (VALUE select key)
Trigger slope Function key
Pre-trigger Function key, Jog (VALUE select key)
RestrictionThe trigger filter cannot be set.
The selected window is not displayed in the pre-
trigger setting.
0V
-600 mV
(2)
200 mV
0V
(1)
NOTE
Settings on the DISPLAY screen
Example for Level Trigger
To cause triggering at point A or point B with the sine wave shown below,
make the following settings.
(1) Point A trigger level: 200 mV, trigger direction (slope): rising (
)
(2) Point B trigger level: -600 mV, trigger direction (slope): falling (
)
When the trigger source is set to AND, the trigger is not enabled when the
trigger slope crosses the set trigger level, but is enabled below or above the
trigger level.