User Manual

270
────────────────────────────────────────────────────
10.7 Trigger Source AND/OR Linking
────────────────────────────────────────────────────
Screen: TRIGGER
Press the TRIG key to call up the
TRIGGER screen.
Move the flashing cursor to the
trigger source item, as shown in the figure
on the left.
Use the function keys to select the AND or
OR.
Link trigger sources with logical
operator OR.
Selection
Selection
Method
Method
Link trigger sources with logical
operator AND.
NOTE
Channe
l
Trigge
r
Trigger leve
l
Slope Filter
CH1 Level 0.00 V OFF
CH2 Level 0.00 V OFF
[
AND linking ]
0V
CH1 waveform
CH2 waveform
CH1 and CH2 level both intersect 0 V line from below
[
OR linking ]
0V
CH1 or CH2 level intersects 0 V line from below
CH1 waveform
CH2 waveform
1
0.7 Trigger Source AND/OR Linking
The analog trigger, logic trigger, external trigger, and timer trigger can be
linked with the AND/OR logical operators.
OR: Triggering occurs when conditions for one trigger are met.
AND: Triggering occurs when conditions for all triggers are met.
If the trigger source is set to AND, and the trigger source setting conditions
have already been met when the START key is pressed, triggering does not
occur. When the conditions are not met once but met subsequently, triggering
occurs.
Example The figures below show the difference between the effect of AND/OR linking.