User Manual
269
────────────────────────────────────────────────────
10.6 Trigger Timing ( REC only )
────────────────────────────────────────────────────
Screen: TRIGGER, DISPLAY
Call up the TRIGGER or DISPLAY screen
.
Move the flashing cursor to the timing
item, as shown in the figure on the left.
Use the function keys to make the
selection.
:Recording starts when the trigger i
s
applied.
Selection
Selection
Method
Method
:Recording starts only when the
START key is pressed, and stops
when the trigger is applied.
:Recording starts when the trigger i
s
applied, and stops when the trigger
is applied next.
NOTE
1
0.6 Trigger Timing ( REC only )
・The moment the trigger occurs, the time related to the waveform being
recorded can be set.
・Not only the waveform after the trigger but also the waveform before the
trigger can be recorded.
When the trigger timing is set to "Stop" or "Start & Stop," measured waveform
data equivalent to the recording length is saved if the stop triggering is not
generated between the initiation of measurement and the end of the set
recording length.
Trigger mode and stop mode:
SINGLE: Stop measurement.
REPEAT (STOP): Start measurement again.
REPEAT (START&STOP):Wait for trigger.