User Manual
241
────────────────────────────────────────────────────
9.13 Setting the 8940 F/V UNIT
────────────────────────────────────────────────────
9
.13.1 Setting Frequency Mode
Screen: CHANNEL1, DISPLAY
Call up the CHANNEL1 or DISPLAY
screen.
Use the CH.SET key to display the desired
channel screen.
Use the CURSOR key to move the flashing
cursor to the graph position.
Use the function keys or the JOG control
to set the FREQ..
:Move the arrow up in the selection
window.
:Move the arrow down in the
selection window.
Selection
Selection
Method
Method
Flashing curso
r
(1) Setting the measurement mode
Flashing curso
r
(2) Setting the measurement range
Move the flashing cursor to the position
shown in the figure on the left.
Use the function keys, the JOG control or
the RANGE key to make the selection.
:Move the arrow up in the selection
window.
:Move the arrow down in the
selection window.
Selection
Selection
NOTE
Select the measurement range. For the Frequency mode, three types of
measurement can be performed depending on the measurement range.
Frequency can be measured from 0.05 Hz to 5 kHz, RPM can be measured
from 5 to 500 rpm, or commercial power can be measured at P50 or P60 Hz.
In the Frequency mode, the measurement range can be set to P50Hz or P60Hz
for commercial power. Although 0 Hz is the usual zero position, in this case 50
or 60 Hz becomes the zero position.